34
3.6.1 - NON RANGEABLE - mV measurement without scaling
Pushing s or t push-button, select the NON RANGEABLE input
type.
The display will show:
Where :
20, 200, 2000 , 20000
show that a fix range is desired.
These values are selectable and have the following significance:
20mV, 200mV, 2000mV, 20000mV f.s.
When a fixed range is selected the instrument will use a fix
decimal figures.
AUTO
indicates that the instrument automatically selects the proper
input range in order to measure the input signal with the max.
possible resolution.
Pushing s or t push-button select the desired input range (e.g.
AUTO)
Push the FUNC push-button.
The instrument starts measuring and the display will show:
Where:
MEAS indicates that the instrument is performing a measurement.
AUTO shows that the autoranging function is selected.
mV is the engineering unit.
XXXXX shows the measured value.
3.6.2 - RNG(L) mV measurement with linear read-out scaling.
The RNG(L) selection allows to associate to an input range the
desired read-out range.
Example 1: with a linear input range from 0 V to 10 V it is desired
to have a read-out going from -100.0 to 1000.0 (mBar).
Pushing s or t push-button, select the RNG (L) function.
Push the FUNC push-button, the display will show the following:
By the numerical keyboard, set the desired initial scale value (For
Example 1, this value is equal to 0 mV).
NOTES:
1) The value assigned to "H.RANGE mV" must be higher than the
value assigned to "L.RANGE mV" parameter.
2) The value assigned to "H.RANGE EU" parameter must be
different from the value assigned to "L.RANGE EU".
20 200 2000
20000 AUTO
MEAS AUTO
XXXXXX mV
L.RNG XXXXXX mV
L.RNG XXXXXX EU
M20-e-3.p65 7/9/01, 2:40 PM34