25
Class II Biological Safety Cabinets
5.4.1.2. Exhaust Filter Replacement for AC2-_D_ and AC2-_G_ Class II BSC
Figure 5.3. Filter Replacement for AC2-_D_ and AC2-_G_
1. Remove the screws from the bottom of blue panel and then lift up the front panel.
Caution: Use the cylindrical struts to carefully hold open the front panel. Continue to work only after the front
panel is secured.
2. Move sash window to the lowest position.
Note: For motorized sash BSC, remove the window straps after moving the sash window to the lowest position.
3. Remove red plenum cover using Phillips screwdriver.
4. Remove the secondary exhaust filter mounting frame from the front.
5. Remove the secondary exhaust filter.
6. Remove the primary exhaust filter from the front.
7. To install the new filter, reverse Steps 1 to 6.
a. When blower housing onto new filter is secured, please ensure gasket compression is equal
on both downflow and exhaust filter.
b. When the exhaust filter mounting frame is secured, please do not over-tighten the nuts, use
the lowest torque if using a power tool.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Esco Class II and is the answer not in the manual?

Esco Class II Specifications

General IconGeneral
LightingLED
UV LightOptional
CertificationNSF/ANSI 49, EN 12469
Airflow PatternLaminar
Inflow Velocity0.38 m/s ± 20%
Downflow Velocity0.25 m/s ± 20%
HEPA Filter Efficiency99.99% at 0.3 μm
MaterialStainless steel
Noise Level< 65 dBA
Front Opening200 mm (8")
ApplicationMicrobiology, cell culture
TypeBiosafety Cabinet

Related product manuals