Model eye
Setting of model eye
• Remove the contact lens holder and set the model eye carefully not to incline back and
forth and around.It cannot take CYL value data correctly if the model eye is inclined.
• Set the model eye at the position where an alignment mark is located at the center of
the reticle mark and the model eye comes into focus.
• When all of the conditions above are met, start the measurement.
6. PERIODICAL INSPECTION AND MAINTENANCE
To prevent malfunction and accidents and maintain to performance and reliability of the product, it is
recommended to request your distributor for the periodical inspection and maintenance once a year.
The periodical inspection and maintenance include inspection of the function and performance of the product,
and cleaning, adjustment and replacement of consumable parts if necessary.
It is recommended for the distributors to perform the cleaning of each part, performance check and accuracy
check at least once in a year.
• Cleaning of each part: exterior parts and optical system.
• Performance check: main unit and each switch.
• Accuracy check: measurement function of refractive power and corneal curvature radius.