3. ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC Standard (IEC 60601-1-2 Ed. 3.0: 2007).
1. This product needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this manual.
2. Portable and mobile RF communications equipment can affect medical electrical equipment.
3. The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the equipment or system as replacement parts for
internal components, may result in increased emissions or decreased immunity of the equipment or
system.
4. The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal operation
in the configuration in which it will be used.
5. The use of the accessory, transducer or cable with equipment and systems other than those specified
may result in increased emission or decreased immunity of the equipment or system.
Guidance and manufacturer's declaration – electromagnetic emissions
AKR550 is intended for use in the electromagnetic environment specified below.
The customer or the user of AKR550 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
AKR550 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class A
AKR550 is suitable for use in all establishments other
than domestic and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
RF emissions CISPR 14-1
RF emissions CISPR 15
USER MANUAL> II. SAFETY CONSIDERATION
11
AKR550 - Auto Kerato Refractometer > V8 - 12-2017

Other manuals for Essilor AKR 550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Essilor AKR 550 and is the answer not in the manual?

Essilor AKR 550 Specifications

General IconGeneral
BrandEssilor
ModelAKR 550
CategoryMedical Equipment
LanguageEnglish

Related product manuals