35 Generator G11
Generator G11 Operator’s Manual
Maintenance and Repair
Periodic calibration is not required for the Generator G11� Periodic check of output using GEN11VK is recommended per facility guidelines�
Service of the Generator G11 would be required if GEN11VK shows the generator is out of tolerance� See GEN11VK Instructions For Use for
guidance on performing the output check� For servicing activities, the Generator G11 may also be returned to an authorized Ethicon service
facility at any time�
The Generator G11 contains a Potential Equalization Terminal on the back panel� This is provided for compatibility with other medical systems
requiring such connections� This conductor is not intended for protective earthing�
The software revision may be displayed on the biomed screen in the user interface�
Disposing of Ethicon Endo-Surgery Generator G11 (Environmental Protection)
The Generator G11 and accessories must not be disposed of at the end of life with other waste� To recycle waste equipment, obtain instructions
from the Ethicon Customer Service Department (call 1-877-ETHICON, +1-513-337-8901 (English) or contact your local sales representative
to discuss local waste disposal solutions and processes� The Generator G11 poses disposal risks similar to consumer electronics such as
computers� There are no radioactive substances, batteries, or hazardous liquids that may leak in the Generator G11�