Generator G11 4
Generator G11 Operator’s Manual
Cart
Figure 3B
General Warnings
Verify that the unit is fully operational prior to administering power output for tissue sealing�
This equipment is for use only by qualified medical personnel trained in the use of ultrasonic surgery or electrosurgery� Inappropriate use
of the equipment by untrained medical personnel may result in hazardous electrical output�
Do not use in the presence of flammable anesthetics or oxidizing gases such as nitrous oxide (N
2
O) and oxygen as combustion may occur
Non-flammable agents should be used for cleaning and disinfection wherever possible� Flammable agents used for cleaning or
disinfecting, or as solvents of adhesives, should be allowed to evaporate before the application of electrosurgery� There is a risk of pooling
of flammable solutions under the patient or in body depressions such as the umbilicus, and in body cavities such as the vagina� Any fluid
pooled in these areas should be mopped up before the electrosurgeryAttention should be called to the danger of ignition of endogenous
or other flammable gases� Some materials, for example cotton and gauze, when saturated with oxygen may be ignited by sparks produced
in normal use during electrosurgery
Do not operate the Generator G11 in a moist environment as a shock hazard may exist� If liquids have entered the Generator G11, the unit
must be returned to the manufacturer for testing prior to use�
Do not operate the unit in close proximity to volatile solvents such as methanol or alcohol as combustion may occur
Avoid use of the Generator G11 adjacent to or stacked with other equipment� If adjacent or stacked use is necessary, monitor the
Generator G11 and the other equipment to assure normal operation�
Interference produced by the operation of high-frequency surgical equipment may adversely affect the operation of other electronic
medical equipment such as monitors and imaging systems�
For patients with cardiac pacemakers or other active implants, a possible hazard exists because interference with the action of the
pacemaker may occur, or the pacemaker may be damaged� In case of doubt, approved qualified advice should be obtained�
Use of accessories and cables other than those specified may result in unpredictable performance, increased electromagnetic emissions, or
decreased electromagnetic immunity
No customer modification of this equipment is allowed; modification of this equipment could have a negative impact on electrical safety
and electromagnetic emissions�
Other than fuses, the Generator G11 contains no operator-serviceable parts� For service, return the generator to an authorized Ethicon
service facility