Generator G11 40
Generator G11 Operator’s Manual
Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1�2 √P
80 MHz to 800 MHz
d = 1�2 √P
800 MHz to 2.5 GHz
d = 2�3 √P
0�01 0�12 0�12 0�23
0�1 0�38 0�38 0�73
1�0 1�20 1�20 2�30
10 3�79 3�79 7�27
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer�
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies�
Note: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people�
Go to www.e-ifu.com for latest revision of service manual�
Footswitch
Non-ionizing Radiation
Equipotential Grounding
Electrical and Electronic equipment� Return waste to a collection system or treatment and recycling facilities�
Applicable in the EU� Follow decontamination instructions before returning waste�
Authorized Representative in the European Community
Authorized Representative in the USA
Manufacturer / Date of Manufacture