170
*Data is from primary sensors only; glucose values between 40 and 400 mg/dL.
Since the two studies are similar, the PRECISION study is included in this report. For detailed information on the
PRECISE II results, please go to our website www.eversensediabetes.com.
Study
Number of
Participants
Total
Number
of Paired
CGM
and YSI
Values
Percent
Within
15/15%
Percent
Within
20/20%
Percent
Within
30/30%
Percent
Within
40/40%
MARD
PRECISE II 90 adults
(18 years
and older)
15,753 87% 94% 99% 100% 8.5%
PRECISION 35 adults
(18 years
and older)
15,170 85% 93% 98% 99% 9.6%
Table 1 – Accuracy to YSI in PRECISE II* and PRECISION*
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 170LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 170 2/26/20 12:59 PM2/26/20 12:59 PM