EasyManua.ls Logo

eversense CGM - Calibration Phases

eversense CGM
213 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
45
Eversense CGM User Guide
Calibration Phases
A. Initialization Phase (after 24-hour Warm-Up Phase)
During this phase, 4 fingerstick blood glucose meter tests are required.
The 4 calibration tests must be spaced 2 to 12 hours apart, and all 4tests must be completed within a 36 hour
period. After 8 hours without a calibration entry, no glucose data will be displayed.
– 1
st
calibration = 24 hours after sensor insertion.
– 2
nd
calibration = 2 to 12 hours after 1st successful calibration.
– 3
rd
calibration = 2 to 12 hours after 2nd successful calibration.
– 4
th
calibration = 2 to 12 hours after 3rd successful calibration.
Glucose readings will start displaying in the app a few minutes after the 2nd calibration is successfully completed.
IMPORTANT: If your smart transmitter is not turned on and paired with the Eversense App and sensor, the
system is not able to prompt you to calibrate.
Re-Entering Initialization Phase
The following will cause the system to re-enter Initialization Phase.
Not completing a calibration test within a 12-hour period during the Initialization Phase.
Not completing all 4 calibration tests within 36 hours during the Initialization Phase.
Not completing 2 calibration tests within a 24-hour period during the Daily Calibration Phase
(see B. Daily Calibration Phase).
When the last several blood glucose meter measurements are significantly dierent than the sensor glucose values.
If the smart transmitter is out of battery power for more than 16 hours.
When you receive a Sensor Check Alert.
Six hours after you receive a Sensor Suspend Alert.
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 45LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 45 2/26/20 12:58 PM2/26/20 12:58 PM

Table of Contents

Related product manuals