Eversense CGM User Guide
B. Daily Calibration Phase
The Daily Calibration Phase requires a blood glucose meter test every 12 hours. The first Daily Calibration Phase will
begin after successful completion of the Initialization Phase.
12 hours after your last successful calibration, the system prompts you to calibrate.
You may optionally enter calibrations more frequently. Daily calibration entries must be spaced at least one hour
apart.
If you do not calibrate within 16 hours, you will receive a Calibration Past Due Alert and no glucose values will be
displayed until a calibration value is entered. After 24 hours without a calibration value entered, you will receive a
Calibration Expired Alert and the system returns to Initialization Phase.
Warning: Certain conditions and alerts will prevent glucose data from being displayed. During these times,
you should use a home blood glucose monitor to make treatment decisions. You should carefully read the
Alerts and Notifications section of this User Guide to understand these conditions.
LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 46LBL-1602-01-001 Rev N_Eversense User Guide_mgdL_R1.indd 46 2/26/20 12:58 PM2/26/20 12:58 PM