B. Daily Calibration Phase
The Daily Calibration Phase requires a blood glucose meter test every 12 hours. The first Daily Calibration Phase will
begin after successful completion of the Initialization Phase.
• 12 hours after your last successful calibration, the system prompts you to calibrate.
• You may optionally enter calibrations more frequently. Daily calibration entries must be spaced at least one hour
apart.
• If you do not calibrate within 16 hours, you will receive a Calibration Past Due Alert and no glucose values will be
displayed until a calibration value is entered. After 24 hours without a calibration value entered, you will receive a
Calibration Expired Alert and the system returns to Initialization Phase.
Warning: Certain conditions and alerts will prevent glucose data from being displayed. During these times,
you should use a home blood glucose monitor to make treatment decisions. You should carefully read the
Alerts and Notifications section of this User Guide to understand these conditions.