19
177
Eversense CGM User Guide
Table 8 – Eversense Percentage of Readings in Agreement on Day 60 in the PRECISION Study
Percent of CGM System Readings Within
CGM
System
Glucose
Range
(mg/dL)
Paired
CGM
and YSI
References
Percent
15/15%
of
Reference
Percent
20/20%
of
Reference
Percent
30/30%
of
Reference
Percent
40/40%
of
Reference
> 40/40%
of
Reference
Overall 2,133 86.9% 93.7% 98.5% 99.6% 0.4%
40 - 60 165 98.2% 100.0% 100.0% 100.0% 0.0%
61 - 80 176 89.2% 96.6% 100.0% 100.0% 0.0%
81 - 180 880 85.9% 94.2% 98.3% 99.4% 0.6%
181 - 300 592 79.1% 86.8% 97.1% 99.5% 0.5%
301 - 350 239 96.7% 100.0% 100.0% 100.0% 0.0%
351 - 400 81 98.8% 100.0% 100.0% 100.0% 0.0%