19
179
Eversense CGM User Guide
Table 10 – Eversense Agreement when “LO” or “HI” is Displayed in the PRECISION Study
Glucose values below 40 mg/dL or above 400 mg/dL are shown on the mobile app as “LO” or “HI”. The table below
shows, as a percentage, agreement of sensor glucose values when outside the display range of 40 mg/dL to 400 mg/dL.
Reference mg/dL
CGM
Readings
CGM-Ref
pairs
< 50 < 60 < 70 < 80 > 80 Total
‘LO’
n 6 7 9 9 0 9
% 66.7% 77.8% 100% 100% 0% 100%
Reference mg/dL
CGM
Readings
CGM-Ref
pairs
> 340 > 320 > 280 > 240 < 240 Total
‘HI’
n 359 383 399 404 1 405
% 88.6% 94.6% 98.5% 99.8% 0.2% 100%