PARD (Paired Absolute Relative Dierence) and Percent Coecient of Variation
In the PRECISION Study, 27 participants had two sensors inserted. The table below shows the precision of the sensor’s
performance by comparing readings between the two sensors worn by each participant. As an example, the absolute
relative dierence between the sensors worn by each person is 9.9%, and the percent coecient of variation is 7.0%,
showing the sensor values generally in agreement throughout the 90 day study period.