BH3 Series
42
EXCELL PRECISION CO., LTD.
The display shows
the maximum capacity as the
calibration weight which could
be changed if
needed.
The calibration value on display
keeps flashing. Put the weight on
the platter. And
press
↵
key.
g
g
4-2 WEIGHT CALIBRATION
Press MODE key to exit the setting and go back the main menu and the display shows
.
.
key ⇒ Upward digit selecting 0,1,2……9
key ⇒ Downward digit selecting 9,8,7……0
key ⇒ Move the cursor one step to the left
key ⇒ Move the cursor one step to the right
↵
key ⇒ Confirm key
Press
↵
key
Press
↵
key
Press
↵
key
After the weight calibration is
finished, take the weight off the
platter. And press
↵
key.
The display shows the maximum capacity
as the calibration weight which could be
changed if needed.
The calibration weight should be no
less than 1/2 of full capacity.