Press key before finishing sampling, it will exit and back to weighing mode.
After Power-off, the scale automatically memorizes the sampling number, and it is
available when “Pcs” unit is selected next time.
When the setting “automatic average unit weight “is turned on, if the object on platter > the
previous sampling number more than 5 pcs, and also <200% the previous sampling
number, the scale will execute unit weight calibration automatically.
2-3 Accumulation
Manual Accumulation
Please set as following: P2 CoM ModE Pr2.
Manual accumulation will sum up the individual weighing values into the memory by
pressing key. This can be repeated 99 times or up to the capacity of the system.
To check accumulation counts and accumulation weight by pressing key.
To print by pressing key.
To clear the accumulation data when there is no weight by pressing + keys. Then
M+ indication “”disappears.
Auto. Accumulation
Please set as following: P2 CoM ModE Auto.
Auto. accumulation will sum up the individual weighing values into the memory
automatically. This can be repeated 99 times or up to the capacity of the system.
To check accumulation counts and accumulation weight by pressing key.
To print by pressing key. To clear the accumulation data when there is no weight by
pressing + keys. Then M+ indication “”disappears.