Colon Insufflator User Manual
Page
14
Section 9.0
A
SSEMBLY PRIOR TO COLON INSUFFLATION
IMPORTANT
If at any time the unit performs erratically or provides otherwise abnormal operation,
remove the unit from service and have it inspected or repaired.
The PROTOCO
2
L Colon Insufflator should be inspected upon receipt and before
each use. Damaged equipment should be removed from service and returned to
E
-
Z
-
EM, Inc. for repair or replacement. Before each use, perform the procedures and
inspections described in Sectio
ns 9.1, 9.2 and 9.3
9.1
INSUFFLATOR PREPARATION
1.
Install the PROTOCO
2
L Colon Insufflator on the Accessory Cart, or on a flat
surface, away from potential sources of spraying or leaking liquids.
2.
Visually inspect the PROTOCO
2
L Colon Insufflator for exter
nal signs of
damage.
9.2
ELECTRICAL CONNECTIONS
1.
Inspect the electrical connections. Do not use if inspection reveals any damage.
2.
Connect the Power Cord to the AC Power Connection on the back of the
PROTOCO
2
L Colon Insufflator.
3.
Before connecting the p
ower cord to the hospital grade wall outlet, make sure
that the main power switch is off and that the voltage is correct. The
PROTOCO
2
L Colon Insufflator has a universal AC line input device, the
nominal AC line voltage is 100 to 240 VAC and the AC line f
requency is
50/60
Hz. The AC line voltage should not drop below 90 VAC or exceed
264 VAC. There are no switches or other AC line configuration requirements.