Section 14.0
TROUBLESHOOTING GUIDE
Make sure that you have read and
understand the prior sections of this operator’s manual that provide
normal operating instructions for your PROTOCO
2
L Colon Insufflator, including the warnings and
cautions section.
Before making arrangements with E
-
Z
-
EM, Inc. Customer Solutions to send your
PROTOCO
2
L Colon Insufflator
unit back to the factory for service, we ask that you take a few minutes
to review
the following information in this guide while simultaneously examining your unit.
Please be advised that the appearance of your
PROTOCO
2
L Colon Insufflator may differ slightly from
the units shown in the following photographs. (Example: buttons may be rou
nd or square).
The Field Checkout steps listed here pertain to common operating conditions and possible malfunctions.
Review this information and refer to it when communicating with your E
-
Z
-
EM, Inc. Customer Solutions
Representative. This will enable us t
o identify the best course of action to meet your service needs.
Should you need further assistance with this guide, see your supervisor, or contact E
-
Z
-
EM, Inc.,
Customer Solutions at 1
-
800
-
544
-
4624 (USA), 1
-
516
-
333
-
8230.
Condition
Possible Causes
Fie
ld
Checkout
My
PROTOCO
2
L Colon
Insufflator
does not
Power
-
up at all.
What should I do?
No power at wall outlet
Blown fuse
Check
PROTOCO
2
L Colon Insufflator
in wall outlet known
to be operational.
Remove power cord from rear panel of
PROTOCO
2
L
C
olon Insufflator
and use a small screwdriver to open fuse
holder.
Pull fuse holder down and examine fuses.
If blown, replace with 0.5
-
Amp 250
-
Volt Time
-
Delay Fuses.