Colon Insufflator User Manual
Page
38
Table 4
Recommended separation distances bet
ween Portable and mobile RF communications
equipment and the PROTOCO
2
L
The PROTOCO
2
L Colon Insufflator is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the PROTOCO
2
L Colon
Insufflator
can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the PROTOCO
2
L Colon Insufflator as
recommended below, according to the maximum output po
wer of the communications equipment.
Separation distance according to frequency of transmitter
m
Read maximum output
power of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
W
d
= 1,2
P
d
= 1,2
P
d
= 2,3
P
0,01
0,12
0,12
0,2
3
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of
the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guid
elines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.