EasyManua.ls Logo

First Stop Safety MemoryPAT-Blu - Downloading to PC; Battery Charging

First Stop Safety MemoryPAT-Blu
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FIRST STOP SAFETY 27
MemoryPAT-Blu Issue 1.2
Once the number of tests carried out exceeds 1990, the tester warns the user
that ALL test results will be erased once more than 2000 test results are in
memory. After this warning has been displayed for the last 10 tests, ALL test
results are automatically erased and the unique test number resets to 1.
DOWNLOADING TO PC
The MemoryPAT-Blu Download software can be used to transfer all the test
information from the tester to a PC.
This Download utility is available on the website www.firststopsafety.co.uk in the
MemoryPAT section. Follow the instructions on the website to download and
install this software and the USB drivers.
First plug in the MemoryPAT-Blu into the USB port of this PC and then launch
the software. There are two large icons available to download or erase test
results. Once the test results have been downloaded to the PC they can easily
be exported to either Excel or CSV format and used for reporting.
BATTERY CHARGING
When the battery capacity is low, a red Battery Low led will come ON, warning
the user to recharge at the next available opportunity. If the battery charge is too
low, the unit will not operate.
The MemoryPAT-Blu is charged using a USB lead plugged
into a wall charger which is shown here.
When the charger is first plugged in a fast charge current of
upto 200mA is applied if the battery voltage is less than 9V.
Once the voltage has reached this level a top-up charge of
70mA is applied for the next 12 hours. During the whole of
the charging period, a green charge led will be ON. (When the charger is first
plugged in, please allow a few seconds for the green led to come on).
If the battery is flat then when the charger is plugged in, the red battery low led
will light followed after some time by the green charging led. After a wait of 5
minutes the MemoryPAT-Blu can be used for testing.