The low temperature alarm will activate an audible/visual warning when
the freezer chamber temperature has reached or decreased below the low
temperature alarm set point. To set the low temperature alarm set point:
1. Press the Mode key until the Set Low Alarm indicator lights.
2. Press the up or down arrow key until the desired low temperature
alarm set point is displayed.
3. Press Enter to save the setting.
4. Press the Mode key until the Run indicator lights for Run mode
If no control keys are pressed, the freezer will automatically return to RUN
mode after 5 minutes.
Note The low alarm set point must be set at least 5°C from the control set
point.
s
Run mode is the default mode for the freezer. This mode will display the
cabinet temperature on the temperature display under normal operating
conditions. In addition, Run mode allows display of the high stage heat
exchange temperature.
This information is scrolled by pressing the up or down arrow keys. The
display will return to the operating temperature in 10 seconds if no keys
are pressed.
1-8 Isotemp Basic Fisher Scientific
Section 1
Installation and Start-Up
Set the Low Temperature
Alarm
Run Mode