EasyManua.ls Logo

Fisher Scientific Isotemp Basic -86C - Section 6 Specifications

Fisher Scientific Isotemp Basic -86C
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 6 Specifications
Isotemp Basic 6-1Fisher Scientific
IE1386A (8955)
-50C to -86C (-58F to -123F) in an 18C to 32C* (64.4F to 89.6F) ambient
33.3”W x 77.9”H x 32.9”
84.6 x 197.9 x 78.7cm
23.0”W x 51.5”H x 19.3”
58.4 x 130.8 x 49.0cm
13.0 cu. ft. (368.1 liters)
Two 1 HP (2545 BTUH each)
Non-CFC, foamed-in-place urethane: 5.0” (12.7 cm) cabinet; 4.5” (11.4 cm) door
120V, 60 Hz, 16.0 FLA
Operating Range:
108VAC-130VAC
20 amp Dedicated Circuit,
20 amp Time Delay Breaker
711 lbs. (323 kg)
IE1386D (8954)
33.3”W x 77.9”H x 32.9”
84.6 x 197.9 x 78.7cm
23.0”W x 51.5”H x 19.3”
58.4 x 130.8 x 49.0cm
13.0 cu. ft. (368.1 liters)
208-230V, 60 Hz, 12.0FLA
Operating Range:
208VAC-240VAC
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
711 lbs. (323 kg)
IE1386V (8953)
33.3”W x 77.9”H x 32.9”
84.6 x 197.9 x 78.7cm
23.0”W x 51.5”H x 19.3”
58.4 x 130.8 x 49.0cm
13.0 cu. ft. (368.1 liters)
230V, 50 Hz, 12.0FLA
Operating Range:
208VAC-240VAC
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
711 lbs. (323 kg)
Catalog No. (Model)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical-
nominal voltage ±10%
Breaker
Requirements
Shipping Weight:
Motor
*Compressors may not cycle off with cabinet running at -86C in a 32C ambient.