EasyManua.ls Logo

Fisher Scientific Isotemp Basic -86C - Connect Probe;Solenoid Harness

Fisher Scientific Isotemp Basic -86C
48 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. Remove the four screws on the freezer back panel and use them to
mount the tie wrap anchors as shown in Figure 5-5. Secure the probe
wire with tie wraps.
2. Plug the solenoid/probe connector into the BUS connection and secure
with a screw on the right and left side. The connector is keyed.
3. Loosen the terminal screws on the solenoid. Slide the spade lug
connectors under the screws and tighten to secure.
4. Connect power to the freezer. Turn the freezer On, with battery switch
Off (O).
a. The Solenoid Engaged light on the BUS control panel will
illuminate (no injection occurs). This light stays on until the unit is
below BUS setpoint.
b. The Low Battery indicator may also illuminate.
5. Turn the battery switch to Standby mode ( ) to charge both batteries.
5-4 Isotemp Basic Fisher Scientific
Section 5
Factory Options
Connect the
Probe/Solenoid Harness
Remote alarm
contacts and
analog output
Power switch
(mains disconnect)
Power Inlet
RS-232 or RS-485
Interface
BUS connection
Spade lug
connections
to solenoid
Tie wrap
anchor
Probe wire
Figure 5-5. Routing

Related product manuals