A Menu Structure FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10 147
Install Medium
selection of the displays for the input of fluids
Install Medium with Special Funct.\SYSTEM settings\
Libraries\Extended Library = off
selection whether a user defined fluid is to be edited or deleted
selection of a user defined fluid
input of a designation for the selected fluid
input of the average sound speed of the fluid
range: 500.0...3500.0 m/s
input of the area around the average sound speed of the fluid
range: 50...999 m/s
input of the kinematic viscosity of the fluid
range: 0.01...30 000.00 mm²/s
input of the operation density of the fluid
Install Medium with Special Funct.\SYSTEM settings\
Libraries\Extended Library = on
selection of the function for the temperature and pressure dependence of the
fluid parameters
selection of a user defined fluid
selection whether the user defined fluid is to be edited or deleted
This display will only be indicated if the selected fluid already exists.
input of a designation for the selected fluid
input of the constants for the longitudinal sound speed of the fluid
The number of constants depends on the function selected above.
input of the kinematic viscosity of the fluid
input of the operation density of the fluid
INIT-resis-
tant
Special Funct. ↕
Install Medium
Install Medium
>EDIT< delete
USER Medium ↕
#01:--not used--
EDIT TEXT (↑↓← →)
USER Medium 1
c-Medium=1500m/s
range +-150m/s
Kinem.Viscosity
1.01 mm2/s
Edit Medium ↕
Basics:Y=m*X +n
USER Medium ↕
#01:--not used--
USER MEDIUM 2
>EDIT< delete
#2: Input Name:
USER MEDIUM 2