FLUXUS F70x A Menu Structure
174 UMFLUXUS_F7V4-6-2EN, 2017-10-01
Input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if Cut-off Flow = sign and Cut-off
Flow = user has been selected.
x
Input of the cut-off flow for the absolute value of the measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if Cut-off Flow = absolut and Cut-off
Flow = user has been selected.
x
Input of the max. signal amplification.
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
x
Input of the fixed upper limit of the sound speed.
range: 0...3 000 m/s
0: the default value 1 848 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
Input of the offset.
range: 0...900 m/s
0: the default value 300 m/s is used
This display will only be indicated if the SuperUser mode is activated.
x
selection of the unit of measurement for the heat quantity x
activation of the transmission and storing of the heat quantity totalizer values
during the heat flow measurement
x
activation of the overflow of the totalizers x
activation of the taking-over of the totalizer values after a restart of the measure-
ment
x
input of the time interval without any valid measured values after which the
transmitter recognizes a long measurement failure
0: the default value 30 s is used
This display will only be indicated if the SuperUser mode is activated.
x
input of the number of decimal places for the totalizers:
Automatic: dynamic adjustment
Fixed to x digit: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
x
activation of the temperature-based heat flow cut-off
This display will only be indicated if the SuperUser mode is activated.
x
INIT-
resistant
-Cut-off Flow
-2.5 cm/s
Cut-off Flow
2.5 cm/s
A: Gain threshold
Fail if > 90 dB
A: Bad soundspeed
thresh. 2007 m/s
A: Bad soundspeed
offset: +321 m/s
Heat Quantity
>[J]< [Wh]
heat+flow quant.
off >ON<
Quant. wrapping
off >ON<
Quantity recall
off >ON<
Do not total. if
no meas.> 0 s
Total digits
Automatic
Thermal low cut
off >ON<

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F706**-A2 Series and is the answer not in the manual?

Flexim FLUXUS F706**-A2 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F706**-A2 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals