FLIR MR160 USER MANUAL Document Identifier: MR160-en-US_AF
20
5.7 High Moisture Alarm
The MR160 offers a High Moisture Alarm where an audible and visual alert activates when the
moisture reading exceeds the programmed high limit.
1. Press the Select button to access the Main Menu
2. Select the SETTINGS mode
from the Main Menu
3. Scroll to ALARM and press the Select button to open the Alarm programmer
4. Use the Navigation and Select buttons to switch the alarm ON or OFF and to set the
threshold from 0% to 100%
5. Press the Select button to return to the SETTINGS mode and save the value or press the
back button to cancel and to return to the normal operating mode
6. When the High Alarm is set ON, the main display will show the alarm bell icon (shown in
header above) and the High Alarm threshold value. Refer to example screens shown in
Fig. 5-1.
7. When the measurement exceeds the threshold, the text for the measurement reading
will appear red in color and will flash. Note that in the Moisture-only mode, the bargraph
turns red when the Alarm threshold is exceeded. Refer to the example in Fig. 5-1(b).
8. To switch the alarm OFF when the meter is alarming, press the Select button and then
select SETTINGS from the Main Menu