Operation
20
The following values are possible:
OFF The heating plate will not be heated (no heat output).
1 – 9 Level of humidifier output, where 1 is the lowest
and 9 is the highest level.
During treatment, the display of the humidifier setting disappears again after
2 sec (no key pressed) and the current therapy display is shown.
We recommend starting with a setting of 3. If further problems (e.g. with a
runny nose) occur, you can increment the setting a point at a time until you
find the optimum setting. At high humidities, problems can arise due to water
condensing in the mask and hose. In this case we recommend the use of a
heated hose system. Please contact your supplier to order this.
5.7 Therapy displays
The FLO Vigaro has 3 different therapy displays.
Therapy display 1
(default)
Current target pressure
Bar display for actual pressure / current pressure
- Current pressure display
Display of current target and actual pressure
+ bar display
Therapy display 2
Tidal volume Breathing rate
Inhalation time Breathing I/E ratio
- Tidal volume
Displays current tidal volume
-
Tidal volume not active or
could not be determined
=
current tidal volume in the range of
tidal volume Vt - tidal volume Vt + 20%