Declarations regarding electromagnetic compatibility
48
19 Declarations regarding electromagnetic compatibility
Guidelines and manufacturer's declaration - Electromagnetic emissions
The FLO Vigaro is intended for use in the electromagnetic environment specified below.
The user of the FLO Vigaro should ensure that it is used in such an environment.
Electromagnetic environment -
Guidelines
RF emissions
as per CISPR 11
The FLO Vigaro uses RF energy only for its
internal function. For this reason its RF
emissions are very low and it is not likely to
cause any interference in nearby electronic
equipment.
RF emissions
as per CISPR 11
The FLO Vigaro is suitable for use in all
establishments including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
as per IEC 61000-3-2
Voltage fluctuations /
flicker emissions as per
IEC 61000-3-3
Guidelines and manufacturer's declaration - Electromagnetic immunity
The FLO Vigaro is intended for use in the electromagnetic environment specified below.
The user of the FLO Vigaro should ensure that it is used in such an environment.
Electromagnetic
environment - Guidelines
Electrostatic
discharge (ESD)
as per
IEC 61000-4-2
± 6 kV
Contact discharge
± 8 kV
air discharge
± 6 kV
Contact discharge
± 8 kV
air discharge
Floors should be made of wood,
concrete or ceramic tiles. If
floors are covered with
synthetic material, the relative
humidity must be at least 30%.