Alarms
34
No.
Display
Error
Continued usage
C25
The alarm limit
for under-
shooting the
tidal volume has
not been met
Yes
C26
The alarm limit
for exceeding
the tidal volume
has been
exceeded
Yes
C28
External voltage
is too low
No (if necessary
recharge the ext.
battery)
9 Help with therapy problems
Problem
Cause
Solution
Noise develops in
the hose
Connections have
become detached
Check that all connections
are secure
Damage to the
breathing hose
Replace the hose, contact
your supplier
Noise develops in
the mask
The mask connection
and the breathing hose
sleeve have become
detached
Check that all connections
are secure
Mask leakage
Contact your supplier