Alarms
8.1 Medium priority alarms
8.1.1 Pressure too high
Alarm
Actual pressure value too high (C30)
Display
Alarm
Actual value exceeds set target value by 20%
Delay
Cause of alarm present for at least 2 seconds
Acoustic alarm
Medium priority
Visible alarm signal
Yellow LED flashes
Possible cause
Hose system blocked
Deactivate
It is not possible to deactivate the alarm.
Comment
The device switches off automatically.
No further operation possible.