Cleaning / Disinfection / Processing
36
10 Cleaning / Disinfection / Processing
The following FLO Vigaro components form part of the breathing gas route and
can be removed by the user / patient:
- Particulate filter:
- Glass
- Air humidifier lid
- Humidifier / equipment connecting joint
These components require regular cleaning and maintenance. This is described
in the following section.
10.1 General
Disconnect the device from the mains power supply
before you start cleaning!
Care and cleaning by the user / patient is limited to cleaning and visual
inspection of the device and mains cable and to replacing device filters, the
hose system and any accessories used.
The user can also disinfect the beaker (see 10.5 Disinfection).
-
Do not submerge the device in water and do
not spray it with water or other liquids
-
Only use mild cleaning products
-
Do not use any Solvents or abrasive cleaning products
-
Cleaning liquids must not be allowed to get inside the device
-
Replace damaged parts
-
After cleaning allow the device to dry off completely
Always carry out cleaning according to the instructions provided by the
manufacturer of the cleaning agents. Cleaning should be carried out on the
basis of applicable hygiene control procedures of the relevant facility.
The FLO Vigaro is not a sterile medical device!