Alarms
30
8.1.2 Pressure too low
Actual pressure value too low (C31)
Actual value is below set target value by 20%
Cause of alarm present for at least 10 seconds
Medium priority
whilst cause of alarm present
- Hose system not properly connected
- Major leakage occurring
The acoustic alarm can be confirmed indefinitely
by pressing the menu key. This deactivation
ceases when the therapy ends or if the cause of
the alarm is no longer present.
The device does not switch off.
As soon as the fault is corrected, the alarm
disappears from the display, and there is no more
acoustic signal.