Declarations regarding electromagnetic compatibility
Electrical fast
transients /
bursts as per
IEC 61000-4-4
± 2 kV for power
lines
± 1 kV for
input and output
lines
± 2 kV for power
lines
± 1 kV for
input and output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage surges as
per IEC 61000-4-5
± 1 kV
line-to-line
± 2 kV
line-to-earth
± 1 kV
line-to-line
voltage
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short interruptions
and voltage
variations on supply
as per
IEC 61000-4-11
< 5% U
T
for 1/2 cycle
(> 95% dip)
40% U
T
for 5 cycles
(60% dip)
70% U
T
for 25 cycles
(30% dip)
< 5% U
T
for 5 s
(> 95% dip)
< 5% U
T
for 1/2 cycle
(> 95% dip)
40% U
T
for 5 cycles
(60% dip)
70% U
T
for 25 cycles
(30% dip)
< 5% U
T
for 5 s
(> 95% dip)
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the FLO Vigaro requires
continued function during
interruptions in the power
supply, it is recommended that
the FLO Vigaro is powered from
an uninterruptible power supply
or battery.
Power frequency
magnetic field
(50/60 Hz)
according to
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be characteristic of
the values found in a typical
commercial or hospital
environment.
Note: U
T
is the AC supply voltage prior to application of the test levels