Technical data
54
Warming the breathed air:
- without the humidifier
- with the humidifier
+1.8 °C
+39.8 °C
Operation
Continuous operation
Operating Range
4 to 30 hPa ( cmH2O)
The maximum pressure < 40 hPa
( cmH2O) is not exceeded even at the
first error
Sound pressure level
19.48 dB (L
pAist,
DIN EN ISO 3744)
Sound power level
< 39 dB(A)
Sound pressure level
Alarm messages
min. 70 db(A)
System resistance with
air flow of 60 l/min at
the patient connection
opening
Minimum stable
pressure limit (PLS min)
0 hPa
(minimum pressure in event of fault)
Maximum stable
pressure limit (PLS max)
40 hPa
(maximum pressure in event of fault)
Dimensions
16.5 x 16.5 x 33.5 cm
Weight
3.9 kg, including warm air humidifier
and integrated power supply
Humidification system
Meets requirements of
DIN EN 60601-1, DIN EN ISO 8185,
Integrated warm air humidifier