Usage and service notes
•
The pressure loading tool is a delicate instrument and
must be handled with care.
•
Use only approved hydraulic oil in the hand pump.
•
It is critical that service technicians be trained in the
assembly and disassembly procedures of the intensi
-
fier before they use the pressure loading tool.
•
Do not use the pressure loading tool for anything
other than the assembly and disassembly of the
intensifier.
•
It is important to collapse the tool (push the piston
fully into its cylinder) before each use as described in
this section, to prevent accidental separation of the
tool.
•
Inspect the tool piston stop for damage. Replace if it
is not flat. DO NOT grind to remove damage.
•
For ease of use and to prevent damage, Flow recom
-
mends that you mount the hand pump permanently
to a rigid cart. When not in use, cover the cart with a
dust cover and store in a clean location.
•
When storing the pressure hose, do not kink or bend
it.
•
Always keep dust caps in place on the quick discon-
nect fittings when the tool is not in use.
•
Once a year, the pressure gauge of the tool should
be calibrated by a qualified laboratory.
18 | M-376 © Flow International Corporation
94K HYPERPRESSURE INTENSIFIER