1736/1738
Users Manual
72
Intrinsic Uncertainty ±(% of measurement value + % of power range)
Parameter
Influence
Quantity
Direct Input
[1]
iFlex1500-12 iFlex3000-24 iFlex6000-36 i40S-EL
Clamp: 50 mV/500 mV
Rogowski: 15 mV/150 mV
150 A/1500 A 300 A/3000 A 600/6000 A 4 A/40 A
Active Power P
Active Energy E
a
PF 0.99 0.5 % + 0.005 % 1.2 % + 0.005 % 1.2 % + 0.0075 % 1.7 % + 0.0075 % 1.2 % + 0.005 %
0.1 PF
<0.99
󰇧
0.5 +
√1 
3×
󰇨
%
+0.005 %
󰇧
1.2 +
√1 
2×
󰇨
%
+ 0.00 5 %
󰇧
1.2 +
√1 
2×
󰇨
%
+ 0.0075 %
󰇧
1.7 +
√1 
2×
󰇨
%
+ 0.0075 %
󰇧
1.2 + 1.7 ×
√1 

󰇨
%
+0.005 %
Apparent Power S
Apparent Energy
E
ap
0 PF 1 0.5 % + 0.005 % 1.2 % + 0.005 % 1.2 % + 0.0075 % 1.7 % + 0.0075 % 1.2 % + 0.005 %
Reactive Power Q
Reactive Energy E
r
0 PF 1 2.5 % of measured apparent power/energy
Power Factor PF
Displacement
Power Factor
DPF/cosφ
- Reading ± 0.025
Additional
uncertainty
(% of power high-
range)
V
P-N
>250 V 0.015 % 0.015 % 0.0225 % 0.0225 % 0.015 %
[1] Only for calibration laboratories
Reference Conditions:
Environmental: 23 °C ±5 °C, instrument operating for at least 30 minutes, no external electrical/magnetic field, RH <65 %
Input conditions: CosΦ/PF=1, Sinusoidal signal f=50/60 Hz, power supply 120 V/230 V ±10 %.
Current and power specifications: Input voltage 1ph: 120 V/230 V or 3ph wye/delta: 230 V/400 V
Input current > 10 % of current range
Primary conductor of clamps or Rogowski coil in center position
Temperature Coefficient: Add 0.1 x specified accuracy for each degree C above 28 °C or below 18 °C