Power Logger
Function Selection Buttons
31
2. Push (Correct Digitally) to access the
connection correction screen. This screen allows you
to virtually swap phases and invert the current inputs
instead of a manual correction.
3. If the Logger is able to determine a better phase map
or polarity, push (Auto Correct) to apply the new
settings.
Auto Correct is not available if the algorithm is not able
to detect a better phase map or when no errors are
detected.
Note
It is impossible to detect all incorrect hook-ups
automatically. You must verify the suggested
modifications carefully before you apply the digital
correction. Applications with single-phase energy
generation can deliver the wrong results when you
apply the Auto Correct feature.
The algorithm works in a way that creates in three phase
systems a sequence with a clockwise phase rotation.
Power
– In the Power mode you can get the values and a
live trend chart for each phase (A, B, C or L1, L2, L3) and
total as:
Active Power (P) in W
Apparent Power (S) in VA
Non-active Power (D) in var
Power Factor (PF)
Use (Fundamental/RMS) to toggle between full
bandwidth power values and power of the fundamental.
In the fundamental power screen you see these values:
Fundamental Active Power (P
fund
+) in W
Fundamental Apparent Power (S
fund
) in VA
Fundamental Reactive Power (Q
fund
) in var
Displacement Power Factor (DPF) / cosϕ
Push (Show Menu) to open a list of simplified Power
screens that show all phases and total of one parameter,
all parameters of one phase, or total.

Table of Contents

Other manuals for Fluke 1738

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1738 and is the answer not in the manual?

Fluke 1738 Specifications

General IconGeneral
Product typeThree phase power measurement logger
Cable lock slot typeKensington
International Protection (IP) codeIP50
Operating altitude0 - 4000 m
Non-operating altitude0 - 12000 m
Storage temperature (T-T)-20 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)10 - 95 %
Built-in displayYes
Display diagonal4.3 \
Touchscreen typeResistive
Display resolution480 x 272 pixels
Display technologyTFT
Efficiency68.2 %
Battery life4 h
Charging time6 h
Battery voltage3.7 V
Input frequency50/60 Hz
Battery capacity9.25 Wh
Power consumption (max)50 VA
Battery life (power save mode)5.5 h
Weight and Dimensions IconWeight and Dimensions
Depth55 mm
Width198 mm
Height167 mm
Weight1100 g
Power adapter weight400 g
Power supply dimensions (WxDxH)130 x 45 x 130 mm

Related product manuals