True-rms Multimeters
Electrical Specifications
21
Function Range
[1]
Resolution
Accuracy ±( [ % of Reading ] + [ Counts ] )
175 177 179
Diode test 2.400 V 0.001 V 1 % + 2
Capacitance
1000 nF
10.00 μF
100.0 μF
9999 μF
[4]
1 nF
0.01 μF
0.1 μF
1 μF
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
1.2 % + 2
1.2 % + 2
1.2 % + 2
10 % typical
AC Amps
[5]
(True-rms)
(45 Hz to 1 kHz)
60.00 mA
400.0 mA
[6]
6.000 A
10.00 A
[7]
0.01 mA
0.1 mA
0.001 A
0.01 A
1.5 % + 3 1.5 % + 3 1.5 % + 3
DC Amps
[5]
60.00 mA
400.0 mA
[6]
6.000 A
10.00 A
[7]
0.01 mA
0.1 mA
0.001 A
0.01 A
1.0 % + 3 1.0 % + 3 1.0 % + 3
Hz
(AC- or DC- coupled,
V or A
[8] [9]
input )
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 1 0.1 % + 1 0.1 % + 1

Other manuals for Fluke 177