175, 177, 179
Users Manual
22
Function Range
[1]
Resolution
Accuracy ±( [ % of Reading ] + [ Counts ] )
175 177 179
Temperature
[10]
-40 °C to +400 °C
-40 °F to +752 °F
0.1 °C
0.1 °F
NA NA
1 % + 10
[11]
1 % + 18
[10]
MIN MAX AVG
For dc functions, accuracy is the specified accuracy of the measurement function ±12 counts for changes longer than 350 ms in
duration.
For ac functions, accuracy is the specified accuracy of the measurement function ±40 counts for changes longer than 900 ms in
duration.
[1] All ac voltage and ac current ranges are specified from 5 % of range to 100 % of range.
[2] Crest factor of 3 at full scale up to 500 V, decreasing linearly to crest factor 1.5 at 1000 V.
[3] For non-sinusoidal waveforms, add -(2 % reading + 2 % full scale) typical, for crest factors up to 3.
[4] In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2 for all models.
[5] Amps input burden voltage (typical): 400 mA input 2 mV/mA, 10 A input 37 mV/A.
[6] 400.0 mA accuracy specified up to 600 mA overload.
[7] >10 A unspecified.
[8] Frequency is specified from 2 Hz to 99.99 kHz in Volts and from 2 Hz to 30 kHz in Amps.
[9] Below 2 Hz, the display shows zero Hz.
[10] In RF field of 3 V/m specified accuracy ±5 °C (9 °F).
[11] Does not include error of the thermocouple probe.