28 II Ex
Getting Started
18
Touch top half
of input contacts
Good 11 A fuse: 0.0 Ω to
0.5 Ω
Replace fuse: OL
Good 0.44 A fuse: 0.995 kΩ to
1.005
Replace fuse: OL
grt08.eps
Figure 2. Current Fuse Test
How to Replace the Batteries
Replace the batteries with three AAA
batteries (NEDA 24A IEC LR03).
XWWarning
To prevent electrical shock or personal
injury:
Replace the batteries when the low
battery indicator () shows to prevent
incorrect measurements. If the display
shows “batt” the Product will not
function until the batteries are replaced.
Use only three AAA 1.5-volt batteries,
correctly installed to power the Product.
See the table on the subsequent page for
a list of approved batteries. All cells are
to be replaced at the same time with
same part number cells in fresh air
locations only.