3130
Users Manual
6
Calibrator Interface
The pressure controls are shown in Figure 3.
V mA
SWITCH TEST
COM
EXTERNAL
PRESSURE MODULE
24 V DC
30 V MAX
LOOP
PORTABLE PRESSURE CALIBRATOR
3130
FLUKE CORPORATION
EVERETT, WA USA
www.flukecal.com
CHARGE
16 V DC
5.6 A
SERIAL NO.
VENT LINE BEFORE MAKING SELECTION
CAUTION
OUTPUT
ISOLATION
VALVE
-12 PSI TO 300 PSI
(-8 kPa TO 2 MPa)
AIR SUPPLY
330 PSI MAX
(2.3 MPa)
PRESSURE
CLOSED
VACUUM
PUMP
FINE
ADJUSTMENT
VENT
SUPPLY
METERING VALVE
FINE
ADJUSTMENT
VENT
FILTER
PRESSURE
SENSOR
CLOSE VALVE BEFORE
OPERATING PUMP
F1 F2 F3
HOME
ZERO PUMP
3
21
4
9
5
6
78
hbb006.eps
Item Description
Supply metering valve. Use the supply metering valve to regulate the incoming air supply.
See the “Typical Connections” section for more.
Isolation valve. This valve isolates the right-most port and pressure sensor from the
Vent/Fine adjust, pump, and Supply Metering Valve.
Pump Pressure/Vacuum. In the + position, the pump operates from 0 psi to 300 psi. In
the - position, the pump runs in vacuum mode, which is 0 psi to -12 psi.
Fine adjust/vent. Use the small silver knob of this valve to bleed off the pressure in a
controlled manner. If left open, it will vent all pressure off safely. The outer-most knob
(black) is a vernier control that varies the volume which finely reduces or increases the
pressure.
Power Adapter input.
24 V power for voltage transmitters. This jack only powered when in Voltage Measure
mode.
Common Jack.
Input Jacks. Use the input jacks for volts and milliamp functions. See the “Measure and
Source Current (4… 20 mA)” and “Measure Voltage” sections. Also used for switch test.
External pressure module connection. See the “External Modules” section.
Figure 3. Controls