Table 6. Error Codes
Error Number Error Description
100 A non-numeric entry was received where it should be a numeric entry
101 Too many digits entered
102 Invalid units or parameter value received
103 Entry is above the upper limit of the allowable range
104 Entry is below the lower limit of the allowable range
105 A required command parameter was missing
106 An invalid command parameter was received
107 Pressure not selected
108 Invalid sensor type
109 Pressure module not connected
110 An unknown command was received
111 Bad Parameter received
112 The serial input buffer overflowed
113 Too many entries in the command line
114 The serial output buffer overflowed
Enter Commands
Commands for the Product can be entered in upper or lower case. There is a minimum of
one space necessary between the command and parameter. All other spaces are optional.
Most of the commands for the Product are sequential. Any overlapped commands will be
shown as such. This section describes each of the commands and their general use, which
will include any parameters that can be entered with the command as well as what the
output of the command is.
Common Commands
*CLS
Clears the error queue. Also ends all pending operations. When you write programs, use
this command before each procedure to prevent buffer overflow.
*IDN?
Shows the manufacturer, model number, and firmware revision of the Product.
For example:
*IDN?
Shows Fluke, 3130 0, 1.00