Calibrate a Pressure-to-Current Transmitter
To calibrate a pressure-to-current transmitter (P/I):
1. Connect the Product pump to the transmitter.
2. Apply pressure from the pump.
3. Measure the current output of the transmitter.
4. Make sure that the read value is correct. If it is not correct, the transmitter must be
adjusted. See Figure 26.