Portable Pressure Calibrator
Pressure Measurement
25
Calibrate a Pressure-to-Voltage Transmitter
To calibrate a pressure-to-voltage transmitter (P/V):
1. Connect the Product pump to the transmitter.
2. Apply pressure from the pump.
3. Connect the 24 V to the transmitter.
4. Measure the voltage output of the transmitter.
5. Make sure that the read value is correct. If it is not correct, the transmitter must be
adjusted. See Figure 27 for connections.
V mA
SWITCH TEST
COM
EXTERNAL
PRESSURE MODULE
24 V DC
30 V MAX
LOOP
PORTABLE PRESSURE CALIBRATOR
3130
FLUKE CORPORATION
EVERETT, WA USA
www.flukecal.com
CHARGE
16 V DC
5.6 A
SERIAL NO.
VENT LINE BEFORE MAKING SELECTION
CAUTION
OUTPUT
ISOLATION
VALV E
-12 PSI TO 300 PSI
(-8 kPa TO 2 MPa)
AIR SUPPLY
330 PSI MAX
(2.3 MPa)
PRESSURE
CLOSED
VACUUM
PUMP
FINE
ADJUSTMENT
VENT
SUPPLY
METERING VALVE
FINE
ADJUSTMENT
VENT
FILTER
PRESSURE
SENSOR
CLOSE VALVE BEFORE
OPERATING PUMP
F1 F2 F3
HOME
ZERO PUMP
Signal Output
Common
+24 V Power
Pressure
Transmitter
hbb035.eps
Figure 27. Calibrate a Pressure-to-Voltage Transmitter