345
Calibration Manual
22
State Key Action Comments
A16 H Screen for A16:
“AMPS 40A CAL”
“APPLY 40A DC”
“Press <RUN> when ready”
Next State: A17
A17 H Screen for A17:
“AMPS 40A CAL”
“A to ASIC V” – Cal value
“A to ASIC A” – Cal value
“Press <RUN> when ready”
Next State: A18
A18 H Screen for A18:
“AMPS 400A CAL”
“APPLY 400A DC”
“Press <RUN> when ready”
Next State: A19
A19 H Screen for A19:
“AMPS 400A CAL”
“A to ASIC V” – Cal value
“A to ASIC A” – Cal value
“Press <RUN> when ready”
Next State: A20
A20 H Screen for A20:
“AMPS 2000A CAL”
“APPLY 1000A DC”
“Press <RUN> when ready”
Next State: A21
A21 H Screen for A21:
“AMPS 2000A CAL”
“A to ASIC V” – Cal value
“A to ASIC A” – Cal value
“Press <RUN> when ready”
Next State: A22
A22 H Screen for A22:
“SAVING AMPS CAL”
On completion:
“Press <RUN> when ready”
Next State: S6