EasyManua.ls Logo

Fluke 374 FC - Maintenance and Cautions

Fluke 374 FC
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
374 FC/375 FC/376 FC and 902 FC
Calibration Manual
6
Frequency via Jaw
Range
375 FC and 376 FC.............................................5.0 Hz to 500.0 Hz
Resolution ...............................................................0.1 Hz
Accuracy..................................................................0.5 % ±5 digits
Trigger Level ...........................................................5 Hz to 10 Hz,
10 A
10 Hz to 100 Hz,
5 A
100 Hz to 500 Hz,
10 A
Frequency via Flexible Current Probe
Range
375 FC and 376 FC.............................................5.0 Hz to 500.0 Hz
Resolution ...............................................................0.1 Hz
Accuracy..................................................................0.5 % ±5 digits
Trigger Level ...........................................................5 Hz to 20 Hz,
25 A
20 Hz to 100 Hz,
20 A
100 Hz to 500 Hz,
25 A
Resistance
Range
374 FC ................................................................6000
Ω
375 FC and 376 FC.............................................60 kΩ
Resolution
374 FC ................................................................0.1
Ω (600.0 Ω)
1
Ω (6000 Ω)
375 FC and 376 FC.............................................0.1
Ω (600.0 Ω)
1
Ω (6000 Ω)
10
Ω (60.00 k Ω)
Accuracy..................................................................1 % ±5 digits
Capacitance
Range......................................................................1000 μF
Resolution ...............................................................0.1
μF (100 μF)
1
μF (1000 μF)
Accuracy..................................................................1 % ±4 digits
902 FC Electrical Specifications
Function Range Resolution Accuracy
Voltage DC 600.0 V 0.1 V 1.0 % ±5 counts
Voltage AC (True-rms) 600.0 V 0.1 V 1.5 % ±5 counts (45 Hz to 400 Hz)
Current AC (True-rms) 600.0 A 0.1 A
2.0 % ±5 counts, (45 Hz to 65 Hz)
2.5 % ±5 counts, (65 Hz to 400 Hz)
Max Crest Factor (50 Hz/60 Hz)
3 @ 180 A
2.5 @ 220 A
1.42 @ 600 A
Note: Add 2 % for C.F. >2
Current DC (True-rms) 200.0
μA0.1 μA 1.0 % ±5 counts
Resistance
600.0
Ω
6000 Ω
60.00 kΩ
0.1 Ω
1 Ω
0.01 kΩ
1.0 % ±5 counts
Continuity <30
Ω -- --
Temperature -40.0 °C to 400.0 °C 0.1 °C 1.0 % ±8 counts
Capacitance
100.0
μF
1000
μF
0.1
μF
1
μF
1.0 % ±4 counts