Clamp Meter
902 FC Electrical Specifications
7
Mechanical Specifications
Size (L x W x H)
374 FC/375 FC/376 FC.......................................249 mm x 85 m x 45 mm
902FC ................................................................. 230 mm x 83.7 mm x 45.4 mm
Weight
374 FC/375 FC/376 FC.......................................410 g
902 FC ................................................................ 382 g
Environmental Specifications
Operating Temperature...........................................-10 °C to +50 °C
Storage Temp
374 FC/375 FC/376 FC....................................... -40 °C to +60 °C
902 FC ................................................................ -30 °C to +60 °C
Operating Humidity ................................................. Non condensing (<10 °C)
≤90 % RH (at 10 °C to 30 °C)
≤75 % RH (at 30 °C to 40 °C)
≤45 % RH (at 40 °C to 50 °C)
Operating Altitude ...................................................2000 meters
Storage Altitude ...................................................... 12 000 meters
Safety
374 FC/375 FC/376 FC.......................................IEC 61010-1, Pollution Degree 2
IEC 61010-2-032: CAT III 1000V / CAT IV 600V
IEC 61010-2-033: CAT III 1000V / CAT IV 600V
902 FC ................................................................ IEC 61010-1, Pollution Degree 2
IEC 61010-2-032: CAT III 600 V / CAT IV 300 V
IEC 61010-2-033: CAT III 600 V / CAT IV 300 V
Intrusion Protection (IP) Rating ..............................IEC 60529: IP30
Radio Frequency Certification................................. FCC ID:T68-FBLE IC:6627A-FBLE
Wireless Radio Frequency Range .......................... 2412 MHz to 2462 MHz
Output Power ......................................................... <100 mW
Electromagnetic Compatibility (EMC)
International ........................................................... IEC 61326-1: Portable, Electromagnetic Environment, IEC 61326-2-2, CISPR 11: Group 1,
Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in
such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may
not meet the immunity requirements of this standard when test leads and/or test probes are connected.
Temperature Coefficients........................................ Add 0.1 x specified accuracy for each °C above 28 °C or below 18 °C