Clamp Meter
Calibration Adjustment
11
Calibration Setup
To set up the Product for calibration:
1. Turn the Product over and use a flat-head screwdriver to remove the battery compartment screw. See Figure 3.
2. Remove the battery door.
3. Remove the batteries.
fig02.eps
Figure 3. Remove Batteries
4. Remove the calibration sticker.
5. Connect the Power Supply to the Product battery terminals.
6. Turn on the Product.
7. Use a small jumper to short the two pads together under the calibration sticker. See Figure 4.
ghn51.eps
Figure 4. Calibration Activation
8. To setup the current calibration see Figure 5.
1
2
3