EasyManua.ls Logo

Fluke 374 FC - Product Handling and Environment

Fluke 374 FC
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
374 FC/375 FC/376 FC and 902 FC
Calibration Manual
4
The Product
XW Warning
To prevent possible electrical shock, fire, or personal injury:
When measuring current with the Jaw, keep fingers behind the Tactile
Barrier
.
fig01.eps
Figure 1. The Product
374 FC/375 FC/376 FC Electrical Specifications
AC Current via Jaw
Range
374 FC and 375 FC.............................................600.0 A
376 FC ................................................................999.9 A
Resolution ............................................................... 0.1 A
Accuracy .................................................................2 % ±5 digits (10 Hz to100 Hz)
2.5 % ±5 digits (100 to 500 Hz)
Crest Factor (50 Hz/60 Hz)
376 FC ................................................................3 @ 500 A
2.5 @ 600 A
1.42 @ 1000 A
374 FC and 375 FC.............................................2.5 @ 350 A
1.42 @ 600 A
Note: Add 2 % for C.F. > 2
AC Current via Flexible Current Probe
Range......................................................................2500 A
Resolution ...............................................................0.1 A (≤999.9 A)
1 A (≤ 2500 A)
Accuracy .................................................................3 % ±5 digits (5 Hz to 500 Hz)
Measurement Category IV is applicable to test and measuring circuits connected at the source of
the building’s low-voltage MAINS installation.
~
This product complies with the WEEE Directive marking requirements. The affixed label
indicates that you must not discard this electrical/electronic product in domestic household
waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I,
this product is classed as category 9 “Monitoring and Control Instrumentation” product. Do not
dispose of this product as unsorted municipal waste.
Table 1. Symbols (cont.)
Symbol Description
Clamp Meter