Remote Display True-rms Clamp Meter
Performance Tests
11
Table 3. Performance Tests
Test
(Switch
Position)
Nominal
Calibrator Output
Value
UUT Meter Reading Limits
Low High
K
AC Volts
20 V 20 V, 20 Hz 19.2 20.8
10 V 10 V, 50 Hz 9.4 10.7
600 V 600 V, 50Hz 590.5 609.5
1000 V 1000 V, 50Hz 980 1020
10 V 10 V, 500 Hz 9.4 10.7
600 V 600 V, 500 Hz 590.5 609.5
1000 V 1000 V, 500 Hz 980 1020
L
DC Volts
-1000 V -1000 V -1015 -985
-600 V -600 V -606.5 -593.5
-10 V -10 V -10.6 -9.4
10 V 10 V 9.4 10.6
600 V 600 V 593.5 606.5
1000 V 1000 V 985 1015
Ohms
0 e 0 e -0.5 0.5
10 e 10 e 9.4 10.6
100 e 100 e 98.5 101.5
590 e 590 e 583.6 596.4
1000 e 1000 e 985 1015
3000 e 3000 e 2965 3035
6000 e 6000 e 5935 6065
10 ke 10 ke 9.85 10.15
30 ke 30 ke 29.65 30.35
60 ke 60 ke 59.35 60.65