Remote Display True-rms Clamp Meter
Calibration Adjustment
17
6. Turn the Rotary Switch to the function you wish to calibrate. If you wish to adjust
Jaw balance, see Balance Adjustment.
7. Push .
8. See Table 5 and apply the required inputs.
9. Push to confirm the value and move the next calibration point.
10. When you are finished with the calibration adjustment, push and then push
to confirm the change.
If you enter the calibration mode but do not wish to save the data, simply turn the Product
OFF.
Balance Adjustment
1. Push to enter the balance calibration mode. The display shows “up”.
2. Place the upper part of the jaw close to coil.
3. Wait for the reading to settle.
4. Push to record the reading. The display shows “don”.
5. Place the lower part of jaw close to coil.
6. Wait for the reading to settle.
7. Push to record the reading.
8. Repeat these steps, until the display shows “End”.
Note
It can take multiple cycles before the display shows “End”.
9. Push to exit balance adjustment.